Skip directly to site content Skip directly to page options Skip directly to A-Z link Skip directly to A-Z link Skip directly to A-Z link
Volume 22, Number 10—October 2016
Dispatch

Reemergence of Mycobacterium chimaera in Heater–Cooler Units despite Intensified Cleaning and Disinfection Protocol

Peter W. Schreiber, Stefan P. Kuster, Barbara Hasse, Cornelia Bayard, Christian Rüegg, Philipp Kohler, Peter M. Keller, Guido V. Bloemberg1, Francesco Maisano, Dominique Bettex, Maximilian Halbe, Rami Sommerstein, and Hugo SaxComments to Author 
Author affiliations: University Hospital Zurich, Zurich, Switzerland (P.W. Schreiber, S.P. Kuster, B. Hasse, C. Bayard, C. Rüegg, P. Kohler, F. Maisano, D. Bettex, M. Halbe, H. Sax); University of Zurich, Zurich (P.W. Schreiber, S.P. Kuster, B. Hasse, C. Bayard, C. Rüegg, P. Kohler, F. Maisano, D. Bettex, H. Sax); National Centre for Mycobacteria at University of Zurich, Zurich (P.M. Keller); Institute of Medical Microbiology at University of Zurich, Zurich (P.M. Keller, G.V. Bloemberg); University Hospital, Bern, Switzerland (R. Sommerstein); University of Bern, Bern (R. Sommerstein)

Main Article

Figure 2

Results of heater–cooler unit (HCU) water surveillance cultures by year and month, University Hospital Zurich, Zurich, Switzerland. The dashed vertical line shows the date of implementation of the intensified protocol (i.e., mid-April 2014). The vertical arrows indicate the start of use of each factory-new HCU in the operating room. HCU 3, HCU 4, and HCU 5 were serviced with the intensified in-house maintenance from the time of delivery. HCU 4 was sent for repair at the manufacturer during Decem

Figure 2. Results of heater–cooler unit (HCU) water surveillance cultures by year and month, University Hospital Zurich, Zurich, Switzerland. The dashed vertical line shows the date of implementation of the intensified protocol (i.e., mid-April 2014). The vertical arrows indicate the start of use of each factory-new HCU in the operating room. HCU 3, HCU 4, and HCU 5 were serviced with the intensified in-house maintenance from the time of delivery. HCU 4 was sent for repair at the manufacturer during December 2014–September 2015. Results of mycobacterial cultures are shown for each HCU. No indicates a negative culture for nontuberculous mycobacteria (NTM), indicated by X. Yes indicates a positive culture for NTM. Filled circles indicate Mycobacterium chimaera; empty circles indicate NTM other than M. chimaera. As per date indicated, cultures were reported as positive if ≥1 water sample (1 sample from cardioplegia and 1 sample from patient circuit were gathered at each date) grew NTM. For missing data points (i.e., no negative or positive results shown), mycobacterial cultures could not be tested because of bacterial overgrowth or lack of availability of the HCU.

Main Article

1Current affiliation: Unilabs, Dübendorf, Switzerland.

Page created: September 20, 2016
Page updated: September 20, 2016
Page reviewed: September 20, 2016
The conclusions, findings, and opinions expressed by authors contributing to this journal do not necessarily reflect the official position of the U.S. Department of Health and Human Services, the Public Health Service, the Centers for Disease Control and Prevention, or the authors' affiliated institutions. Use of trade names is for identification only and does not imply endorsement by any of the groups named above.
file_external