Suboptimal Handling of Piccolo Samples or Reagent Discs for Consideration in Ebola Response

Author affiliations: Centers for Disease Control and Prevention, Atlanta, Georgia, USA

Cite This Article

Abstract

Operating clinical analyzers within recommended parameters can be challenging during outbreak response. Using the Piccolo Xpress point-of-care blood chemistry analyzer on guinea pig blood, we found that values of many analytes are still readily comparable when samples and reagent discs are handled at various conditions outside of manufacturer recommendations.

Blood chemistry analyses are useful for guiding patient care. However, following manufacturer-recommended handling and storage conditions can be challenging in areas with underdeveloped infrastructure, as experienced in past and ongoing Ebola outbreak response (1). To investigate the utility of data from samples or reagent discs handled under suboptimal conditions, we evaluated 14 conditions outside of manufacturer recommendations by using Strain 13/N guinea pig blood and plasma samples. Animal procedures were approved by the Centers for Disease Control and Prevention Institutional Animal Care and Use Committee and conducted at an AAALAC-International–accredited facility.

Samples were run on the Abaxis Piccolo Xpress Chemistry Analyzer (https://www.abaxis.com; quality control with Abbot General Chemistry controls and verification sample, https://www.fishersci.com/shop/products/pic-lpd-pls-gen-chm-ct-2x6x1ml/07p0401a). This platform is a compact and portable Clinical Laboratory Improvement Amendments–approved automated point-of-care system for whole blood, serum, and plasma (2). This platform, together with the General Chemistry 13 reagent disc used here, is widely used in past and ongoing Ebola outbreak responses (36) and in laboratory research on viral pathogenesis, therapeutics, and vaccine efficacy (79). All samples were collected in the recommended lithium heparin (LiH) tubes, except as indicated.

We determined intrinsic variation of each analyte under recommended conditions by running 31 samples on 2 different machines simultaneously or on 1 machine sequentially (represented as mean percentage change ± SD). We then evaluated 2 blood collection anticoagulants, 8 sample storage conditions, and 4 reagent disc storage conditions. Overall, >200 samples from 93 guinea pigs (48 males, 45 females; 7 were <1 month of age, 24 were 1–3 months, 8 were 4–6 months, 16 were 7–12 months, 17 were 1–2 years, 11 were 2–3 years, and 10 were >3 years), from healthy animals or animals with noninfectious chronic disease (e.g., renal failure) were analyzed.

We first evaluated the effect of anticoagulants on sample analytes. Blood was collected and then partitioned from the syringe into paired samples stored in LiH (baseline), EDTA, or sodium citrate. Deviations from baseline were determined and expressed as mean percentage differences (Table; Appendix [expanded color version]). As expected, because of the chelating action of EDTA, assay reactions involving cationic substrates (Ca2+, alkaline phosphatase) generated no values or were greatly altered, although values of several analytes remained closely comparable to baseline. Values from samples collected in sodium citrate were uniformly lower than those collected in LiH.

To assess effects of sample storage temperatures on data output, we conducted similar analyses on whole blood stored overnight at −20°C, −4°C, or room temperature (≈20°C–22°C) and on plasma stored overnight at −20°C, room temperature, or 32°C. In general, plasma was more resistant to suboptimal temperatures; values for most analytes were comparable with paired control. Although data from whole blood samples varied more, many analytes (creatinine, alanine aminotransferase, alkaline phosphatase, total bilirubin) remained within ± 1 SD of baseline values. This finding is useful in situations where centrifuging blood samples might not be possible. For application to high-containment studies, we also evaluated the utility of data from plasma stored overnight at −20°C and γ-irradiated (5 × 106 rads) before analysis. Under these conditions, 6 of 12 analytes varied by >1 SD from expected range, suggesting that samples should be processed before inactivation by γ-irradiation.

Finally, we assessed effects of storing reagent discs at room temperature or 32°C for varying lengths of time. Although discs must be refrigerated when stored >48 hours, all values obtained from discs stored at room temperature for a week were within ± 1 SD of controls, except glucose and γ-glutamyltransferase (± 1–2 SD). Data from discs stored at 32°C for 5 days were similarly comparable with baseline; we observed deviations in glucose, γ-glutamyltransferase, and creatinine. Discs stored at 32°C for 14 days were unreliable; many generated no values or a “disc error” message. However, values read from discs that did not result in an error message were all within expected limits. All discs stored at 32°C for 5 weeks generated error messages. These findings suggest that although extended storage at elevated temperatures substantially damages discs, discs stored at these conditions for up to 2 weeks might still yield clinically relevant data.

In summary, we found that under various suboptimal conditions, many analytes are still readily comparable (within ± 1 SD of intrinsic variation) to those from paired samples handled according to the manufacturer’s recommendations. In addition, the pattern of values that deviate from baseline is often consistent (e.g., glucose decreases) and may be considered in clinical evaluation. Although we used SDs to highlight deviation here, the clinical implication of these findings will vary based on the analyte and condition being monitored. We believe these findings, based on guinea pig samples, enabling us to efficiently sample a large population and process under controlled conditions, are translatable to other species, including humans, because the equivalent veterinary point-of-care platform (VetScan VS2) is also disc-based and uses comparable assay chemistry. However, the possibility for some species differences remains and should be considered for future investigations.

Dr. Spengler is a veterinary medical officer with the Viral Special Pathogens Branch, Division of High-Consequence Pathogens and Pathology, National Center for Emerging and Zoonotic Infectious Diseases, CDC. Her research interests include model development and translational research on pathogenesis and therapeutic interventions for diseases caused by hemorrhagic fever viruses.

Top

Acknowledgments

We thank the staff of the Comparative Medicine Branch, Division of Scientific Resources, National Center for Emerging and Zoonotic Infectious Diseases, Centers for Disease Control and Prevention, for assistance with colony husbandry and maintenance, and Tatyana Klimova for assistance in editing the manuscript.

This study was supported in part by the CDC Emerging Infectious Disease Research Core funds.

Top

References

  1. Carazo Perez  S, Folkesson  E, Anglaret  X, Beavogui  A-H, Berbain  E, Camara  A-M, et al. Challenges in preparing and implementing a clinical trial at field level in an Ebola emergency: A case study in Guinea, West Africa. PLoS Negl Trop Dis. 2017;11:e0005545. DOIPubMedGoogle Scholar
  2. Murata  K, Glaser  L, Nardiello  M, Richardson  S, Ramanathan  LV, Carlow  DC. Analytical performance of the Abaxis Piccolo Xpress® point of care analyzer in whole blood, serum, and plasma. Clin Biochem. 2015;48:13446. DOIPubMedGoogle Scholar
  3. de Wit  E, Rosenke  K, Fischer  RJ, Marzi  A, Prescott  J, Bushmaker  T, et al. Ebola laboratory response at the Eternal Love Winning Africa campus, Monrovia, Liberia, 2014–2015. J Infect Dis. 2016;214(suppl 3):S16976. DOIPubMedGoogle Scholar
  4. de Wit  E, Kramer  S, Prescott  J, Rosenke  K, Falzarano  D, Marzi  A, et al. Clinical chemistry of patients with Ebola in Monrovia, Liberia. J Infect Dis. 2016;214(suppl 3):S3037. DOIPubMedGoogle Scholar
  5. Sissoko  D, Laouenan  C, Folkesson  E, M’Lebing  A-B, Beavogui  A-H, Baize  S, et al.; JIKI Study Group. Experimental treatment with favipiravir for Ebola virus disease (the JIKI trial): a historically controlled, single-arm proof-of-concept trial in Guinea. PLoS Med. 2016;13:e1001967. 2 Errata in: PLoS Med. 2016.
  6. Hunt  L, Gupta-Wright  A, Simms  V, Tamba  F, Knott  V, Tamba  K, et al. Clinical presentation, biochemical, and haematological parameters and their association with outcome in patients with Ebola virus disease: an observational cohort study. Lancet Infect Dis. 2015;15:12929. DOIPubMedGoogle Scholar
  7. Bradfute  SB, Swanson  PE, Smith  MA, Watanabe  E, McDunn  JE, Hotchkiss  RS, et al. Mechanisms and consequences of ebolavirus-induced lymphocyte apoptosis. J Immunol. 2010;184:32735. DOIPubMedGoogle Scholar
  8. Geisbert  TW, Mire  CE, Geisbert  JB, Chan  Y-P, Agans  KN, Feldmann  F, et al. Therapeutic treatment of Nipah virus infection in nonhuman primates with a neutralizing human monoclonal antibody. Sci Transl Med. 2014;6:242ra82.
  9. Geisbert  TW, Jones  S, Fritz  EA, Shurtleff  AC, Geisbert  JB, Liebscher  R, et al. Development of a new vaccine for the prevention of Lassa fever. PLoS Med. 2005;2:e183. DOIPubMedGoogle Scholar
  10. Genzer  SC, Huynh  T, Coleman-McCray  JD, Harmon  JR, Welch  SR, Spengler  JR. Reference intervals for hematology and clinical chemistry of strain 13/N guinea pigs (Cavia porcellus). J Am Assoc Lab Anim Sci. In press.

Top

Table

Top

Cite This Article

DOI: 10.3201/eid2506.181928

Original Publication Date: March 22, 2019

Table of Contents – Volume 25, Number 6—June 2019

EID Search Options
presentation_01 Advanced Article Search – Search articles by author and/or keyword.
presentation_01 Articles by Country Search – Search articles by the topic country.
presentation_01 Article Type Search – Search articles by article type and issue.

Top

Comments

Please use the form below to submit correspondence to the authors or contact them at the following address:

Jessica R. Spengler, Centers for Disease Control and Prevention, 1600 Clifton Rd NE, Mailstop H18-SB, Atlanta, GA 30329-4027, USA

Send To

10000 character(s) remaining.

Top

Page created: May 20, 2019
Page updated: May 20, 2019
Page reviewed: May 20, 2019
The conclusions, findings, and opinions expressed by authors contributing to this journal do not necessarily reflect the official position of the U.S. Department of Health and Human Services, the Public Health Service, the Centers for Disease Control and Prevention, or the authors' affiliated institutions. Use of trade names is for identification only and does not imply endorsement by any of the groups named above.
file_external